Abstract

The results of the Immunochromatographic Method for Infectious Disease Testing (IMLTD) may vary depending on the presence of lipids in donor blood samples, potentially leading to false‐negative or false‐positive outcomes in hepatitis B surface antigen (HBsAg) screening. This study aimed to evaluate the effect of lipid interference on HBsAg screening using the chemiluminescent immunoassay (CLIA) method, to determine the range of triglyceride levels that may influence test results (100 mg/dL to >3000 mg/dL), and to assess practical lipid removal methods to ensure blood safety. An experimental laboratory study was conducted using two sample groups. The first group consisted of six bags of non-lipemic, HBsAg-reactive plasma that were rendered lipemic by the addition of Lipofundin. The second group included 25 bags of lipemic donor plasma that were non-reactive in the initial IMLTD screening and subsequently subjected to lipid removal treatments. Lipid reduction was performed using centrifugation, polyethylene glycol (PEG), and diethyl ether. The results showed no significant differences in HBsAg values before and after Lipofundin addition, indicating that lipemia did not affect HBsAg detection by the CLIA method. However, lipid removal using centrifugation, PEG, and diethyl ether significantly reduced triglyceride and total cholesterol levels (p < 0.05) in lipemic samples that interfered with CLIA analysis. In conclusion, lipemia does not directly affect HBsAg screening results using the CLIA method; however, lipid removal procedures are effective and necessary for managing highly lipemic samples to maintain the accuracy and reliability of blood screening tests.

Downloads

Download data is not yet available.

References

Arulvijayavani, S., Mohanraj, P. S., & Reeta, R. (2023). Evaluating interference of lipemia on routine clinical biochemical tests. Journal of Laboratory Physicians, 15(02), 269-275. https://doi.org/10.1055/s-0042-1758664

Berman, M., & Hayes, J. (2024). A-119 Lipemia Interference. Clinical Chemistry, 70(Supplement_1), 106-118. https://doi.org/10.1093/clinchem/hvae106.118

Burdino, E., Ruggiero, T., Proietti, A., Milia, M. G., Olivero, A., Caviglia, G. P., ... & Ghisetti, V. (2014). Quantification of hepatitis B surface antigen with the novel DiaSorin LIAISON XL Murex HBsAg Quant: Correlation with the ARCHITECT quantitative assays. Journal of Clinical Virology, 60(4), 341-346. https://doi.org/10.1016/j.jcv.2014.05.013

Castro-Castro, M. J., Candás-Estébanez, B., Esteban-Salán, M., Calmarza, P., Arrobas-Velilla, T., Romero-Román, C., ... & Aguilar-Doreste, J. Á. (2018). Removing lipemia in serum/plasma samples: a multicenter study. Annals of laboratory medicine, 38(6), 518. https://doi.org/10.3343/alm.2018.38.6.518

Dean, C. L., Wade, J., & Roback, J. D. (2018). Transfusion-transmitted infections: an update on product screening, diagnostic techniques, and the path ahead. Journal of Clinical Microbiology, 56(7), 10-1128. https://doi.org/10.1128/JCM.00352-18

Dimeski, G., & Jones, B. W. (2011). Effective process for lipid reduction using high speed centrifugation compared with ultracentrifugation. Biochemia Medica, 21(1), 86-92. https://hrcak.srce.hr/64505

Fernández Prendes, C., Castro Castro, M. J., Sánchez Navarro, L., Rapún Mas, L., Morales Indiano, C., & Arrobas Velilla, T. (2023). Handling of lipemic samples in the clinical laboratory. Advances in Laboratory Medicine/Avances en Medicina de Laboratorio, 4(1), 5-15. https://doi.org/doi:10.1515/almed-2023-0003

Fu, P., Devare, S., Liu, J. F., Lv, X. T., Yin, P., Wu, B. T., ... & Shan, H. (2020). Evaluate performance of the Abbott chemiluminescent microparticle immunoassay assay for detection of syphilis infection in Chinese blood donors. Journal of Clinical Laboratory Analysis, 34(1), e23033. https://doi.org/10.1002/jcla.23033

Gardiner, C., Lane, P., Tailor, H., & Mackie, I. J. (2020). A practical method for reducing the interference due to lipaemia in coagulation tests. International Journal of Laboratory Hematology, 42(2), 140-144. https://doi.org/10.1111/ijlh.13129

Ghazal, K., Brabant, S., Prie, D., & Piketty, M. L. (2022). Hormone immunoassay interference: a 2021 update. Annals of laboratory medicine, 42(1), 3-23. https://doi.org/10.3343/alm.2022.42.1.3

Gupta, A., Sheedy, J., Ronzoni, S., & Lin, Y. (2022). Gross lipemia and blood bank testing. Transfusion, 62(12). https://doi.org/10.1111/trf.17103

Hippy, N. S. I., & Ukma, N. (2021). Prevalensi Kasus Reaktif HBsAg Pada Pendonor Dengan Metode ChLIA Dan Elisa Berdasarkan Usia Dan Jenis Kelamin Di Udd Pmi Kota Depok Pada Tahun 2019-2020. Ensiklopedia of Journal, 3(5), 114-120. https://doi.org/10.33559/eoj.v4i3.203

Jacob, E. A., Owoseni, M. F., Ibijola, A., Omisakin, C. T., Omoniyi, D. P., Oyedele, T., & Kolawole, F. (2023). Performance of enzyme linked immunosorbent assay and rapid screening techniques for detection of transfusion transmitted infections among blood donors. African Journal of Laboratory Haematology and Transfusion Science, 2(4), 239-246. https://doi.org/10.59708/ajlhts.v2i4.2326

Karin, A., Higgins, V., Miller, J., Brinc, D., Kulasingam, V., & Selvaratnam, R. (2023). Evaluation of hemolysis, lipemia, and icterus interference with common clinical immunoassays. Clinical Chemistry and Laboratory Medicine (CCLM), 61(6), 1035-1045. https://doi.org/10.1515/cclm-2022-0924

Kisniarti, L., Sukmana, D. J., Aini, A., Alfisahrin, A., & Halid, I. (2023). Prevalensi HBsAg Positif Pada Pendonor Darah Di Unit Daerah Palang Merah Indonesia Lombok Barat. Jurnal Sains Natural, 1(2), 51-55. https://doi.org/10.35746/jsn.v1i2.347

Maksymyuk, H., Boykiv, N., & Martianova, O. (2023). Factors Of Interference In Laboratory Investigations: Hemolysis, Icterus, Lipemia. Ukrainian Journal of Laboratory Medicine, 1(1), 22-32. https://doi.org/10.62151/2786-9288.1.1.2023.04

Martín-Calderón, J. L. (2022). Evaluation of endogenous interferences on several electrochemiluminescent immunoassays. La Rivista Italiana della Medicina di Laboratorio, 18(4), 196-202. https://doi.org/10.23736/S1825-859X.23.00167-6

Negash, M., Ayalew, M., Geremew, D., & Workineh, M. (2019). Seroprevalence and associated risk factors for HIV, Hepatitis B and C among blood Donors in South Gondar District blood Bank, Northwest Ethiopia. BMC infectious diseases, 19(1), 430. https://doi.org/10.1186/s12879-019-4051-y

Pliego-Sandoval, J. E., Díaz-Barbosa, A., Reyes-Nava, L. A., Angeles Camacho-Ruiz, M., Iñiguez-Muñoz, L. E., & Pinto-Pérez, O. (2023). Development and evaluation of a low-cost triglyceride quantification enzymatic biosensor using an Arduino-based microfluidic system. Biosensors, 13(8), 826. https://doi.org/10.3390/bios13080826

Soh, S. X., Loh, T. P., Sethi, S. K., & Ong, L. (2022). Methods to reduce lipemic interference in clinical chemistry tests: a systematic review and recommendations. Clinical Chemistry and Laboratory Medicine (CCLM), 60(2), 152-161. https://doi.org/doi:10.1515/cclm- 2021-0979

Soh, S. X., Loh, T. P., Sethi, S. K., & Ong, L. (2022). Methods to reduce lipemic interference in clinical chemistry tests: a systematic review and recommendations. Clinical Chemistry and Laboratory Medicine (CCLM), 60(2), 152-161. https://doi.org/10.1515/cclm-2021-0979

Winter, K. M., Webb, R. G., & Marks, D. C. (2022). Red cells manufactured from lipaemic whole blood donations: Do they have higher haemolysis?. Vox Sanguinis, 117(12), 1351-1359. https://doi.org/10.1111/vox.13366

Wiranto, R. R. P., Momuat, L. I., & Suryanto, E. (2024). Penggunaan Polietilena Glikol 6000 sebagai Flokulan Lipoprotein pada Serum Lipemik dalam Analisis Trigliserida. Chemistry Progress, 17(1), 97–103. https://doi.org/10.35799/cp.17.1.2024.54823

Downloads

Published

2025-12-31

How to Cite

Astriani , R. D., Ritchie, N. K., & Sadikin, M. (2025). The Effect of Triglycerides and Total Cholesterol Levels of Lipemic Plasma on The Results of HBsAg Screening Using Chemiluminescence Immunoassay. JURNAL INFO KESEHATAN, 23(4), 794–804. https://doi.org/10.31965/infokes.Vol23.Iss4.1809